基于火焰光度检测器的多组分硫化物同步检测技术在化工副产食品级二氧化碳分析中的应用

李嘉皓1,由佳泰1,赵姝婷1,董鑫1,李琬菁1,向淑梅2,万纯1   

  1. 1. 浙江石油化工有限公司
    2. 浙江德荣化工有限公司
  • 收稿日期:2026-06-29 修回日期:2026-08-17 出版日期:2026-12-12 发布日期:2026-11-17
  • 通讯作者: 万纯 E-mail:zsh_wanc@rong-sheng.com

APPLICATION OF FLAME PHOTOMETRIC DETECTOR-BASED MULTI-COMPONENT SULFIDE SIMULTANEOUS DETECTION TECHNOLOGY IN THE ANALYSIS OF FOOD-GRADE CARBON DIOXIDE FROM CHEMICAL BY-PRODUCTS

  • Received:2026-06-29 Revised:2026-08-17 Online:2026-12-12 Published:2026-11-17

摘要: 在化工副产食品级二氧化碳中痕量硫化物含量检测方法中,现行国家标准检测方法需二次进样测定二氧化硫与硫含量,检测效率偏低。本研究建立了气相色谱-火焰光度检测器联用检测方法,实现了食品级二氧化碳中二氧化硫与硫含量的单次进样同步检测。结合多种检测器定性样品中的硫化物组分,对比色谱柱分离性能,优化色谱检测条件,确定最优参数为:J&W Select Low Sulfur专用硫色谱柱,载气为氮气,柱流速为0.7 mL/min,柱箱温度为50 ℃,检测器燃烧池温度为250 ℃,氢气与空气体积比为1:10。方法验证结果表明,所建方法目标硫化物分离度达标,重复性、检出限、准确度均符合《食品安全国家标准 食品添加剂 二氧化碳》(GB 1886.228—2016)标准要求,检测成本低、分析速度快,可用于食品级二氧化碳中硫含量的快速质控检测。

关键词: 气相色谱法, 痕量硫化物, 火焰光度检测器, 同步检测技术

Abstract: Trace sulfide measurement is an essential quality control item for food-grade carbon dioxide recycled from chemical by-products. The existing official national standard testing procedure requires two independent sample injections to quantify sulfur dioxide and total sulfur separately, which significantly reduces testing throughput and prolongs analytical cycle for mass samples. In this work, a gas chromatography coupled with flame photometric detector (GC-FPD) analytical method was developed to realize one single injection for synchronous quantitative analysis of sulfur dioxide and total sulfur contained in food-grade carbon dioxide. Multiple detectors were applied to distinguish and confirm various sulfide species qualitatively. Different sulfur-dedicated chromatographic columns were evaluated in terms of peak separation capacity. A full set of instrumental chromatographic parameters was systematically optimized. The finalized optimal operating conditions are listed as below: a J&W Select Low Sulfur inert chromatographic column, nitrogen as carrier gas, column flow rate of 0.7 mL/min, isothermal oven temperature maintained at 50℃, FPD combustion cell temperature set at 250℃, and a hydrogen-to-air volumetric ratio of 1:10. Comprehensive method validation experiments were carried out to assess analytical performance. The results demonstrate that all target sulfide compounds achieve satisfactory chromatographic resolution. Key performance indicators including repeatability, limit of detection and analytical accuracy fully satisfy the specifications stipulated in Chinese national standard GB 1886.228—2016. Compared with the traditional two-injection standard method, the proposed method cuts down consumption of test gas and auxiliary reagents with lower overall testing cost, and greatly shortens single-sample analysis time. This established GC-FPD method is suitable for rapid on-site sulfur index quality inspection and routine batch quality monitoring of food-grade carbon dioxide from chemical by-product sources.

Key words: gas chromatography, trace sulfides, flame photometric detector, single-injection simultaneous detection