高温气相色谱法测定费-托合成蜡的碳数分布

池朗珠,李自运,蒋莹,吕蝶   

  1. 亚申科技研发中心(上海)有限公司
  • 收稿日期:2015-08-25 修回日期:2015-10-19 出版日期:2016-04-12 发布日期:2016-04-19
  • 通讯作者: 李自运 E-mail:ziyun_li@yashentech.com

DETERMINATION OF CARBON NUMBER DISTRIBUTION OF FISCHER-TROPSCH WAX BY HIGH TEMPERATURE GAS CHROMATOGRAPHY

  • Received:2015-08-25 Revised:2015-10-19 Online:2016-04-12 Published:2016-04-19

摘要: 首次采用柱上进样技术,使用DB-HT-SimDis色谱柱,通过高温气相色谱-火焰离子化检测法,对含高碳数产物的费托合成蜡的碳数分布进行了快速准确的测定。对载气流量、柱箱升温程序、进样量、进样浓度等实验参数进行了全面的考察。在最佳实验条件下,可在60 min内准确测定费托合成蜡C21~C101的碳数分布,所得C21~C74相邻碳数化合物分离度高于1.5,C74~C101之间相邻碳数化合物分离度高于1.2。从C57至C101,Ln(W/n)与对应碳数的线性关系良好,相关系数为0.999 3,可计算得到实验用费托合成产物的链增长因子α值为0.928。所建方法操作简便、分离效率高、测定结果准确,对各种蜡产物的品质分析有重要的意义。

关键词: 高温气相色谱, 火焰离子化检测法, DB-HT-SimDis色谱柱, 费-托合成蜡, 碳数分布

Abstract: A rapid and efficient method was developed to determine the carbon number distribution of Fischer-Tropsch wax by high temperature gas chromatography coupled with cold-on-column inlet and flame ionization detector (FID). The experimental parameters, such as column flow, oven temperature program, injection volume and sample concentration, were studied in detailed by using a DB-HT-SimDis column. The optimized GC conditions were as follows: Column: DB-HT-SimDis 5 m×0.53 mm×0.15 μm. Carrier gas: hydrogen at flow-rate of 9.0 mL/min (constant flow); Flame ionization detector (FID) temperature 400 ℃. Combustion gases: hydrogen, 40 mL/min; air, 400 mL/min. Nitrogen as makeup gas, 30 mL/min; Oven temperature program: initial oven temperature 60 ℃, hold 1 min; ramp 1 rate, 10 ℃/min to 240 ℃, hold 1 min; ramp 2 rate, 5 ℃/min to 400 ℃, hold 10 min; Inlet temperature: track oven temperature; Injection volume, 0.5 μL; Sample concentration in heated n-heptane: 0.5%. Under above conditions, the carbon number distribution of Fischer-Tropsch wax between C21 and C101 was analyzed within 60 minutes accurately. The GC resolutions higher than 1.5 and 1.2 were obtained in the ranges of C21 to C74 and C74 to C101, respectively. The linear correlation coefficient value between Ln(W/n) and carbon number in the range of C57 and C101 was 0.999 3 and the chain growth probability (α value) of 0.928 was calculated.

Key words: high temperature gas chromatography, flame ionization detector, DB-HT-SimDis column, Fischer-Tropsch wax, carbon number distribution